Appendix
0: Maintenance Model DMM7510 7½ Digit Graphical Sampling Multimeter
Reference Manual
2 DMM7510-901-01 Rev. B / May 2015
5. Use the screwdriver to lift the tab from the fuse drawer.
6. Slide the fuse drawer out. The fuse drawer does not pull completely out of the power module.
7. Snap the fuse out of the drawer.
8. Replace the fuse.
9. Push the fuse drawer back into the module.
If the fuse continues to become damaged, a circuit malfunction exists and must be corrected. Return
the instrument to Keithley Instruments for repair.
Input fuse replacement
The input line from the AMPS connectors on the front and rear panels is protected by two fuses on
the rear panel.
Make sure the instrument is disconnected from the power line and other equipment before
checking or replacing a current-input fuse. Failure to disconnect all power may expose you
to hazardous voltages, that, if contacted, could cause personal injury or death. Use
appropriate safety precautions when working with hazardous voltages.
For continued protection against fire or instrument damage, only replace fuse with the type and
rating listed. If the instrument repeatedly damages fuses, locate and correct the cause of the problem
before replacing the fuse.
To replace a current-input fuse:
1. Turn off the power to the instrument.
2. Disconnect the power line and test leads.
3. From the rear panel, gently push in the AMPS fuse holder and rotate it one-quarter turn counter-
clockwise.
4. Remove the fuse, and replace it with the same type (see table below).
5. Install the new fuse by reversing the procedure above.
Manufacturer and part
number
Rating Length
DMM7510-FUSE-3A
3.5 A, 1000 VAC/VDC fast blow
10 mm x 38 mm (0.394" x 1.5")
DMM7510-FUSE-10A
11 A, 1000 VAC/VDC fast blow
10.3 mm x 38 mm (0.406" x 1.5")
If the fuse continues to become damaged, a circuit malfunction exists and must be corrected. Return
the instrument to Keithley Instruments for repair.

Table of Contents

Other manuals for Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM7510 and is the answer not in the manual?

Keithley DMM7510 Specifications

General IconGeneral
BrandKeithley
ModelDMM7510
CategoryMultimeter
LanguageEnglish

Related product manuals