6: SCPI command reference Model DMM7510 7½ Digit Graphical Sampling Multimeter
6-180 DMM7510-901-01 Rev. B / May 2015
:TRIGger:BLENder<n>:STIMulus<m>
This command specifies the events that trigger the blender.
Type Affected by Where saved Default value
Instrument reset
Power cycle
Usage
:TRIGger:BLENder<n>:STIMulus<m> <event>
:TRIGger:BLENder<n>:STIMulus<m>?
The blender number (1 or 2)
The stimulus input number (1 to 4)
Details
There are four stimulus inputs that can each select a different event.
Use zero to disable the blender input.
The <event> parameter may be any of the trigger events shown in the following table.
Trigger events
Event description Event constant
No trigger event
Front-panel TRIGGER key press
Notify trigger block <n> (1 to 8); the trigger model generates a
trigger event when it executes the notify block
A command interface trigger (bus trigger):
• Any remote interface: *TRG
• GPIB only: GET bus command
• VXI-11: VXI-11 command
Line edge (either rising, falling, or either based on the
configuration of the line) detected on digital input line
(1 to 6)
Line edge detected on TSP-Link synchronization line <n> (1 to 3)
Appropriate LXI trigger packet is received on LAN trigger object
<n>
(1 to 8)
Trigger event blender <n> (1 or 2), which combines trigger events
Trigger timer <n> (1 to 4) expired