Section
6: SCPI command reference Model DMM7510 7½ Digit Graphical Sampling Multimeter
Reference Manual
6-198 DMM7510-901-01 Rev. B / May 2015
:TRIGger:BLOCk:LIST?
This command returns the settings for all trigger model blocks.
Type Affected by Where saved Default value
Query only Not applicable Not applicable Not applicable
Usage
:TRIGger:BLOCk:LIST?
Details
This returns the settings for the trigger model.
Example
:TRIG:BLOC:LIST?
Returns the settings for the trigger model. Example output is:
1) BUFFER_CLEAR BUFFER: defbuffer1
2) MEASURE BUFFER: defbuffer1 COUNT: 1
3) BRANCH_COUNTER VALUE: 5 BRANCH_BLOCK: 2
4) DELAY_CONSTANT DELAY: 1.000000000
5) BRANCH_COUNTER VALUE: 3 BRANCH_BLOCK: 2
Also see
None
:TRIGger:BLOCk:LOG:EVENt
This command allows you to log an event in the event log when the trigger model is running.
Type Affected by Where saved Default value
Command only
Recall settings
Instrument reset
Power cycle
Save settings
Not applicable
Usage
:TRIGger:BLOCk:LOG:EVENt <blockNumber>, <eventNumber>, "<message>"
<blockNumber>
The sequence of the block in the trigger model
<eventNumber>
The event number:
• INFO<n>
• WARNing<n>
• ERRor<n>
Where <n> is 1 to 4; you can define up to four of each type
You can also set ABORt, which aborts the trigger model immediately and posts a
warning event log message
<message>
A string up to 31 characters

Table of Contents

Other manuals for Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM7510 and is the answer not in the manual?

Keithley DMM7510 Specifications

General IconGeneral
BrandKeithley
ModelDMM7510
CategoryMultimeter
LanguageEnglish

Related product manuals