Model DMM7510
7½ Digit Graphical Sampling Multimeter Reference Manual Section 8:
TSP command reference
DMM7510-901-01 Rev. B / May 2015 8-321
trigger.model.setblock() — trigger.BLOCK_MEASURE
This function defines a trigger block that makes a measurement.
Type TSP-Link accessible Affected by Where saved Default value
Function
Yes
Restore configuration
Instrument reset
Power cycle
Configuration script
Not applicable
Usage
trigger.model.setblock(blockNumber, trigger.BLOCK_MEASURE)
trigger.model.setblock(blockNumber, trigger.BLOCK_MEASURE, bufferName)
trigger.model.setblock(blockNumber, trigger.BLOCK_MEASURE, bufferName, count)
blockNumber
The sequence of the block in the trigger model
bufferName
The name of the buffer, which must be an existing buffer; if no buffer is defined,
defbuffer1
is used
count
The number of digitize readings to make before moving to the next block in the trigger
model; set to a specific value or infinite (trigger.COUNT_INFINITE) or stop infinite
(trigger.COUNT_STOP)
Details
When trigger model execution reaches the block:
1. The instrument begins making a measurement.
2. The trigger model execution waits for the measurement to complete.
3. The instrument places the measurement into the specified reading buffer, which cannot be of the
writable buffer style.
If you are defining a user-defined reading buffer, you must create it before you define this block.
When you set the count to a finite value, trigger model execution remains at the block until all
measurements are complete. If you set the count to infinite, the trigger model executes subsequent
blocks and measurements continue in the background until the trigger model execution reaches
another measure block or until the trigger model ends.
You must select a measure function before running a trigger model that contains this block.
You cannot include a measure block and a digitize block in the same trigger model.

Table of Contents

Other manuals for Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM7510 and is the answer not in the manual?

Keithley DMM7510 Specifications

General IconGeneral
BrandKeithley
ModelDMM7510
CategoryMultimeter
LanguageEnglish

Related product manuals