7½ Digit Graphical Sampling Multimeter Reference Manual Section 6:
DMM7510-901-01 Rev. B / May 2015 6-159
:TRACe:LOG:STATe
This command indicates if information events are logged when the specified reading buffer is at 0 % or 100 %
filled.
Type Affected by Where saved Default value
Instrument reset
Power cycle
defbuffer2: 1 (ON)
User-created buffer: 0 (OFF)
Usage
:TRACe:LOG:STATe <logState>
:TRACe:LOG:STATe <logState>, "<bufferName>"
:TRACe:LOG:STATe?
:TRACe:LOG:STATe? "<bufferName>"
Do not log information events:
or
Log information events: ON or 1
A string that indicates the reading buffer; the default buffers (defbuffer1 or
defbuffer2) or the name of a user-defined buffer; if no buffer is specified,
Details
If this is set to on, when the reading buffer is cleared (0 % filled) or full (100 % filled), an event is
logged in the event log. If this is set to off, reading buffer status is not reported in the event log.
Example
Query the log state of defbuffer1.
Output:
1
Indicates that the log state is on.
Also see
Reading buffers (on page 3-13)
Remote buffer operation (on page 3-30)
:TRACe:MAKE (on page 6-160)
Using the event log (on page 2-154)