Model DMM7510
7½ Digit Graphical Sampling Multimeter Reference Manual Section 6:
SCPI command reference
DMM7510-901-01 Rev. B / May 2015 6-159
:TRACe:LOG:STATe
This command indicates if information events are logged when the specified reading buffer is at 0 % or 100 %
filled.
Type Affected by Where saved Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
defbuffer1: 1 (ON)
defbuffer2: 1 (ON)
User-created buffer: 0 (OFF)
Usage
:TRACe:LOG:STATe <logState>
:TRACe:LOG:STATe <logState>, "<bufferName>"
:TRACe:LOG:STATe?
:TRACe:LOG:STATe? "<bufferName>"
<logState>
Do not log information events:
OFF
or
0
Log information events: ON or 1
<bufferName>
A string that indicates the reading buffer; the default buffers (defbuffer1 or
defbuffer2) or the name of a user-defined buffer; if no buffer is specified,
defbuffer1 is used
Details
If this is set to on, when the reading buffer is cleared (0 % filled) or full (100 % filled), an event is
logged in the event log. If this is set to off, reading buffer status is not reported in the event log.
Example
TRACe:LOG:STATe?
Query the log state of defbuffer1.
Output:
1
Indicates that the log state is on.
Also see
Reading buffers (on page 3-13)
Remote buffer operation (on page 3-30)
:TRACe:MAKE (on page 6-160)
Using the event log (on page 2-154)

Table of Contents

Other manuals for Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM7510 and is the answer not in the manual?

Keithley DMM7510 Specifications

General IconGeneral
BrandKeithley
ModelDMM7510
CategoryMultimeter
LanguageEnglish

Related product manuals