Model DMM7510
7½ Digit Graphical Sampling Multimeter Reference Manual Section 6:
SCPI command reference
DMM7510-901-01 Rev. B / May 2015 6-9
The output of :MEASure:DIGitize? is affected by the data format selected by :FORMat[:DATA].
If you set FORMat[:DATA] to REAL or SREAL, you will have fewer options for buffer elements. The
only buffer elements available are READing, RELative, and EXTRa. If you request a buffer element
that is not permitted for the selected data format, the instrument generates the error 1133, "Parameter
4, Syntax error, expected valid name parameters."
The STATus buffer element returns status values for the readings in the buffer. The status values are
floating-point numbers that encode the status value. Refer to the following table for values.
Buffer status bits for sense measurements
Bit (hex) Name Decimal Description
0x0001
STAT_QUESTIONABLE
1
Measure status questionable
0x0006
STAT_ORIGIN
6
A/D converter from which reading originated; for the
Model DMM7510, this will always be 0 (Main) or 2
(digitizer)
0x0008
STAT_TERMINAL
8
Measure terminal, front is 1, rear is 0
0x0010
STAT_LIMIT2_LOW
16
Measure status limit 2 low
0x0020
STAT_LIMIT2_HIGH
32
Measure status limit 2 high
0x0040
STAT_LIMIT1_LOW
64
Measure status limit 1 low
0x0080
STAT_LIMIT1_HIGH
128
Measure status limit 1 high
0x0100
STAT_START_GROUP
256
First reading in a group
Example
TRACe:MAKE "voltDigitizeBuffer", 10000
MEAS:DIG:VOLT? "voltDigitizeBuffer", FORM, DATE, READ
Create a buffer named voltMeasBuffer. Make a digitize voltage reading and store it in the buffer
voltMeasBuffer and return the formatted reading, the date, and the reading elements from the buffer.
Example output:
-00.0024 mV,05/16/2014,-2.384862E-06
Also see
:READ:DIGitize? (on page 6-12)
:TRACe:DATA? (on page 6-155)
:READ?
This query makes measurements, places them in a reading buffer, and returns the last reading.
Type Affected by Where saved Default value
Query only
Not applicable
Not applicable
Not applicable
Usage
:READ?
:READ? "<bufferName>"
:READ? "<bufferName>", <bufferElements>
<bufferName>
The name of the buffer where the reading is stored; if nothing is specified,
defbuffer1
is used
<bufferElements>
See Details; if nothing is specified, READing is used

Table of Contents

Other manuals for Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM7510 and is the answer not in the manual?

Keithley DMM7510 Specifications

General IconGeneral
BrandKeithley
ModelDMM7510
CategoryMultimeter
LanguageEnglish

Related product manuals