Model DMM7510
7½ Digit Graphical Sampling Multimeter Reference Manual Section 8:
TSP command reference
DMM7510-901-01 Rev. B / May 2015 8-259
trigger.clear()
This function clears any pending command triggers.
Type TSP-Link accessible
Affected by Where saved Default value
Function
Yes
Usage
trigger.clear()
Details
A command trigger indicates if a trigger event has been detected over a command interface since the
last trigger.wait() command was sent. Command triggers are generated by:
• Sending *TRG over a remote interface
• GET bus commands
• VXI-11 device trigger commands
trigger.clear() clears the command triggers and discards the history of trigger events.
Example
*TRG
print(trigger.wait(1))
trigger.clear()
print(trigger.wait(1))
Generate a trigger event.
Check if there are any pending trigger
events.
Output: true
Clear any pending command triggers.
Check if there are any pending trigger
events.
Output:
false
Also see
trigger.wait() (on page 8-345)

Table of Contents

Other manuals for Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM7510 and is the answer not in the manual?

Keithley DMM7510 Specifications

General IconGeneral
BrandKeithley
ModelDMM7510
CategoryMultimeter
LanguageEnglish

Related product manuals