8: TSP command reference Model DMM7510 7½ Digit Graphical Sampling Multimeter
8-260 DMM7510-901-01 Rev. B / May 2015
trigger.digin[N].clear()
This function clears the trigger event on a digital input line.
Type TSP-Link accessible
Affected by Where saved Default value
Usage
trigger.digin[N].clear()
Digital I/O trigger line (1 to 6)
Details
The event detector of a trigger enters the detected state when an event is detected. For the specified
trigger line, this command clears the event detector, discards the history, and clears the overrun
status(sets the overrun status to false).
Example
Clears the trigger event detector on I/O line 2.
Also see
digio.line[N].mode (on page 8-52)
Digital I/O port configuration (on page 3-49)
trigger.digin[N].overrun (on page 8-261)
trigger.digin[N].wait() (on page 8-262)
trigger.digin[N].edge
This attribute sets the edge used by the trigger event detector on the given trigger line.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW) Yes
Instrument reset
Configuration script trigger.EDGE_FALLING
Usage
detectedEdge = trigger.digin[N].edge
trigger.digin[N].edge = detectedEdge
• Detect falling-edge triggers as inputs: trigger.EDGE_FALLING
• Detect rising-edge triggers as inputs: trigger.EDGE_RISING
• Detect either falling or rising-edge triggers as inputs:
trigger.EDGE_EITHER
•
See Details for descriptions of values
Digital I/O trigger line (1 to 6)