8: TSP command reference Model DMM7510 7½ Digit Graphical Sampling Multimeter
8-136 DMM7510-901-01 Rev. B / May 2015
dmm.measure.autozero.once()
This function causes the instrument to refresh the reference and zero measurements once.
Type TSP-Link accessible
Affected by Where saved Default value
Usage
dmm.measure.autozero.once()
Functions
dmm.FUNC_DIGITIZE_CURRENT
dmm.FUNC_DIGITIZE_VOLTAGE
Details
This command forces a refresh of the reference and zero measurements that are used for the present
aperture setting for the selected function.
When autozero is set to off, the instrument may gradually drift out of specification. To minimize the
drift, you can send the once command to make a reference and zero measurement immediately
before a test sequence.
If the NPLC setting is less than 0.2 PLC, sending autozero once can result in delay of more than a
second.
Example
dmm.measure.autozero.once()
Do a one-time refresh of the reference and zero
measurements.
Also see
Automatic reference measurements (on page 2-149)
dmm.measure.autozero.enable (on page 8-135)