6: SCPI command reference Model DMM7510 7½ Digit Graphical Sampling Multimeter
6-88 DMM7510-901-01 Rev. B / May 2015
[:SENSe[1]]:<function>:ODETector
This command determines if the detection of open leads is enabled or disabled.
Type Affected by Where saved Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
4W Res: OFF (0)
Temperature: ON (1)
Usage
[:SENSe[1]]:<function>:ODETector <state>
[:SENSe[1]]:<function>:ODETector?
The function to which the setting applies; see Functions
Disable:
or
Functions
Details
For temperature measurements, this is only available when the transducer is set to a thermocouple or
one of the RTDs.
Long lengths of thermocouple wire can have a large amount of capacitance, which is seen at the
input of the DMM. If an intermittent open occurs in the thermocouple circuit, the capacitance can
cause an erroneous on-scale reading. The open thermocouple detection circuit, when enabled,
applies a 100 µA pulse of current to the thermocouple before the start of each temperature
measurement.
Example
TEMP:TC:TYPE K
TEMP:UNIT C
Set the transducer type to thermocouple.
Set the thermocouple type to K.
Set the units to Celsius.
Turn open lead detection off.
Also see
None