8: TSP command reference Model DMM7510 7½ Digit Graphical Sampling Multimeter
8-266 DMM7510-901-01 Rev. B / May 2015
trigger.digout[N].stimulus
This attribute selects the event that causes a trigger to be asserted on the digital output line.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW) Yes
Restore configuration
Instrument reset
Power cycle
Digital I/O trigger N reset
Configuration script trigger.EVENT_NONE
Usage
event = trigger.digout[N].stimulus
trigger.digout[N].stimulus = event
The event to use as a stimulus; see Details
Digital I/O trigger line (1 to 6)
Details
The digital trigger pulsewidth command determines how long the trigger is asserted.
The trigger stimulus for a digital I/O line can be set to one of the trigger events that are described in
the following table.
Trigger events
Event description Event constant
Front-panel TRIGGER key press
Notify trigger block N (1 to 8) generates a trigger event when
the trigger model executes it
A command interface trigger (bus trigger):
• Any remote interface: *TRG
• GPIB only: GET bus command
• VXI-11: VXI-11 command
Line edge (either rising, falling, or either based on the
configuration of the line) detected on digital input line N
Line edge detected on TSP-Link synchronization line N
(1 to 3)
Appropriate LXI trigger packet is received on LAN trigger
object N (1 to 8)