6: SCPI command reference Model DMM7510 7½ Digit Graphical Sampling Multimeter
6-204 DMM7510-901-01 Rev. B / May 2015
:TRIGger:DIGital<n>:IN:CLEar
This command clears the trigger event on a digital input line.
Type Affected by Where saved Default value
Command only Not applicable Not applicable Not applicable
Usage
:TRIGger:DIGital<n>:IN:CLEar
Digital I/O trigger line (1 to 6)
Details
The event detector of a trigger enters the detected state when an event is detected. For the specified
trigger line, this command clears the event detector, discards the history, and clears the overrun
status(sets the overrun status to 0).
For this block to work as expected, make sure you configure the trigger type and line state of the
digital line for use with the trigger model (use the digital line mode command).
Example
Clears the trigger event detector on I/O line 2.
Also see
:DIGital:LINE<n>:MODE (on page 6-38)
Digital I/O port configuration (on page 3-49)
:TRIGger:DIGital<n>:IN:OVERrun? (on page 6-205)
:TRIGger:DIGital<n>:IN:EDGE
This command sets the edge used by the trigger event detector on the given trigger line.
Type Affected by Where saved Default value
Instrument reset
Power cycle
Usage
:TRIGger:DIGital<n>:IN:EDGE <detectedEdge>
:TRIGger:DIGital<n>:IN:EDGE?
Digital I/O trigger line (1 to 6)
• Detect falling-edge triggers as inputs: FALLing
• Detect rising-edge triggers as inputs: RISing
• Detect either falling or rising-edge triggers as inputs: EITHer
See Details for descriptions of values