Model DMM7510
7½ Digit Graphical Sampling Multimeter Reference Manual Section 3:
Functions and features
DMM7510-901-01 Rev. B / May 2015 3-117
Using TSP-Net with any ethernet-enabled instrument
Refer to TSP command reference (on page 8-1) for details about the commands presented in this
section.
The Model DMM7510 LAN port is auto-sensing (Auto-MDIX), so you can use either a LAN crossover
cable or a LAN straight-through cable to connect directly from the Model DMM7510 to an ethernet
device or to a hub.
To set up communication to a remote ethernet-enabled instrument that is TSP
®
enabled:
1. Send the following command to configure TSP-Net to send an abort command when a
connection to a TSP instrument is established:
tspnet.tsp.abortonconnect = 1
If the scripts are allowed to run, the connection is made, but the remote instrument may be busy.
2. Send the command:
connectionID = tspnet.connect(ipAddress)
Where:
• connectionID is the connection ID that will be used as a handle in all other TSP-Net function calls.
• ipAddress is the IP address of the remote instrument.
See tspnet.connect() (on page 8-354) for additional detail.
To set up communication to a remote ethernet-enabled device that is not TSP enabled:
Send the command:
connectionID = tspnet.connect(ipAddress, portNumber, initString)
Where:
• connectionID is the connection ID that will be used as a handle in all other tspnet function calls.
• ipAddress is the IP address of the remote device.
• portNumber is the port number of the remote device.
• initString is the initialization string that is to be sent to ipAddress.
See tspnet.connect() (on page 8-354) for additional detail.
To communicate to a remote ethernet device from the Model DMM7510:
1. Connect to the remote device using one of the above procedures. If the Model DMM7510 cannot
make a connection to the remote device, it generates a timeout event. Use tspnet.timeout to
set the timeout value. The default timeout value is 20 s.
2. Use tspnet.write() or tspnet.execute() to send strings to a remote device. If you use:
ï‚§ tspnet.write(): Strings are sent to the device exactly as indicated, and you must supply
any needed termination characters.
ï‚§ tspnet.execute(): The Model DMM7510 appends termination characters to all strings
that are sent. Use tspnet.termination() to specify the termination character.
3. To retrieve responses from the remote instrument, use tspnet.read(). The Model DMM7510
suspends operation until the remote device responds or a timeout event is generated. To check if
data is available from the remote instrument, use tspnet.readavailable().
4. Disconnect from the remote device using the tspnet.disconnect() function. Terminate all
remote connections using tspnet.reset().

Table of Contents

Other manuals for Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM7510 and is the answer not in the manual?

Keithley DMM7510 Specifications

General IconGeneral
BrandKeithley
ModelDMM7510
CategoryMultimeter
LanguageEnglish

Related product manuals