7½ Digit Graphical Sampling Multimeter Reference Manual Section 6:
DMM7510-901-01 Rev. B / May 2015 6-69
Example 1
CURR:AVER:TCON MOV
Set the averaging filter type to moving
average, with a filter count of 10.
Enable the averaging filter.
Example 2
RES:AVER:TCON MOV
Set the averaging filter type to moving
average, with a filter count of 10.
Enable the averaging filter.
Example 3
VOLT:AVER:TCON MOV
Set the averaging filter type to moving
average, with a filter count of 10.
Enable the averaging filter.
Also see
Filtering measurement data (on page 3-11)
[:SENSe[1]]:<function>:AVERage:COUNt (on page 6-67)
[:SENSe[1]]:<function>:AVERage:TCONtrol (on page 6-69)
[:SENSe[1]]:<function>:AVERage:WINDow (on page 6-71)
[:SENSe[1]]:<function>:AVERage:TCONtrol
This command sets the type of averaging filter that is used for the selected measure function when the
measurement filter is enabled.
Type Affected by Where saved Default value
Instrument reset
Power cycle
Measure configuration list
Measure configuration list
Usage
[:SENSe[1]]:<function>:AVERage:TCONtrol <type>
[:SENSe[1]]:<function>:AVERage:TCONtrol?
The function to which the setting applies; see Functions
The filter type to use when filtering is enabled; set to one of the following values:
• Repeating filter: REPeat
•
Functions