7½ Digit Graphical Sampling Multimeter Reference Manual Section 6:
DMM7510-901-01 Rev. B / May 2015 6-149
:SYSTem:TEMPerature:INTernal?
This command returns the internal temperature of the instrument.
Type Affected by Where saved Default value
Query only Not applicable Not applicable Not applicable
Usage
:SYSTem:TEMPerature:INTernal?
Details
Returns the last recorded value of internal temperature of the instrument in Celsius (°C). The
instrument checks internal temperature when it updates references when autozero is on. Internal
temperature is not checked if autozero is set to off. It can also become stale when digitize
measurements are used or when measurements take a long time.
If the temperature changes more than ±5 °C, the instrument logs an event and displays a message
on the front panel that recommends that you perform auto calibration.
Example
Returns the internal temperature of the instrument.
Example output:
Also see
:ACAL:LASTrun:TEMPerature:INTernal? (on page 6-16)
:ACAL:RUN (on page 6-20)