Page #375 background imageLoading...
Page #375 background image
Section
6: SCPI command reference Model DMM7510 7½ Digit Graphical Sampling Multimeter
Reference Manual
6-48 DMM7510-901-01 Rev. B / May 2015
FORMat subsystem
The commands for this subsystem select the data format that is used to transfer instrument readings
over the remote interface.
:FORMat:ASCii:PRECision
This command sets the precision (number of digits) for all numbers returned in the ASCII format.
Type Affected by Where saved Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
0
Usage
:FORMat:ASCii:PRECision <n>
:FORMat:ASCii:PRECision DEFault
:FORMat:ASCii:PRECision MINimum
:FORMat:ASCii:PRECision MAXimum
:FORMat:ASCii:PRECision?
:FORMat:ASCii:PRECision? DEFault
:FORMat:ASCii:PRECision? MINimum
:FORMat:ASCii:PRECision? MAXimum
<n>
The precision:
• Automatic: 0
•
Specific value: 1 to 16
Details
This attribute specifies the precision (number of digits) for queries.
Note that the precision is the number of significant digits. There is always one digit to the left of the
decimal point; be sure to include this digit when setting the precision.
Example
:FORM:ASC:PREC 10
Set a precision of 10 digits. An example of the
output is:
-6.999999881E-01
Also see
:FORMat[:DATA] (on page 6-50)

Table of Contents

Other manuals for Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM7510 and is the answer not in the manual?

Keithley DMM7510 Specifications

General IconGeneral
BrandKeithley
ModelDMM7510
CategoryMultimeter
LanguageEnglish

Related product manuals