7½ Digit Graphical Sampling Multimeter Reference Manual Section 8:
TSP command reference
DMM7510-901-01 Rev. B / May 2015 8-299
trigger.model.load() — SimpleLoop
This function loads a predefined trigger model configuration that does a specific number of measurements.
Type TSP-Link accessible
Affected by Where saved Default value
Function
Yes
Usage
trigger.model.load("SimpleLoop", count)
trigger.model.load("SimpleLoop", count, delay)
trigger.model.load("SimpleLoop", count, delay, bufferName)
trigger.model.load("SimpleLoop", count, delay, bufferName, readingBlock)
count
The number of measurements the instrument will make
delay
The delay time before each measurement (167 ns to 10 ks); default is 0 for
no delay
bufferName
A string that indicates the reading buffer; the default buffers (defbuffer1
or defbuffer2) or the name of a user-defined buffer; if no buffer is
specified, defbuffer1 is used
readingBlock
Define a measure or digitize block for the trigger model; options are:
• trigger.READING_ACTIVE: Add a measure or digitize block to the
trigger model based on the active function; if no option defined,
trigger.READING_ACTIVE is used
• trigger.READING_MEASURE: Adds a measure block to the trigger
model
• trigger.READING_DIGITIZE: Adds a digitize block to the trigger
model
Details
This command sets up a loop that sets a delay, makes a measurement, and then repeats the loop the
number of times you define in the count parameter.
The out line of the EXT TRIG IN/OUT rear-panel terminal is asserted at the end of each
measurement.
After selecting a trigger model template, you can view the trigger model blocks in a graphical format
by pressing the front-panel MENU key and under Trigger, selecting Configure. You can also add or
delete blocks and change trigger model settings from this screen. You can use the
trigger.model.getblocklist() command to view the trigger model blocks in a list format.

Table of Contents

Other manuals for Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM7510 and is the answer not in the manual?

Keithley DMM7510 Specifications

General IconGeneral
BrandKeithley
ModelDMM7510
CategoryMultimeter
LanguageEnglish

Related product manuals