8: TSP command reference Model DMM7510 7½ Digit Graphical Sampling Multimeter
8-298 DMM7510-901-01 Rev. B / May 2015
Trigger events
Event description Event constant
Trigger event blender N (1 to 2), which combines trigger
events
Trigger timer N (1 to 4) expired
trigger.EVENT_ANALOGTRIGGER
External in trigger
After selecting a trigger model template, you can view the trigger model blocks in a graphical format
by pressing the front-panel MENU key and under Trigger, selecting Configure. You can also add or
delete blocks and change trigger model settings from this screen. You can use the
trigger.model.getblocklist() command to view the trigger model blocks in a list format.
The following usage has been deprecated; replace it with the usage above that includes the clear
parameter.
trigger.model.load("LoopUntilEvent", triggerEvent, position)
trigger.model.load("LoopUntilEvent", triggerEvent, position, delay)
trigger.model.load("LoopUntilEvent", triggerEvent, position, delay, bufferName)
trigger.model.load("LoopUntilEvent", triggerEvent, position, delay, bufferName,
readingBlock)
Example
-- Set up measure function
dmm.measure.func = dmm.FUNC_DC_CURRENT
-- Initiate readings
trigger.model.load("LoopUntilEvent", trigger.EVENT_DISPLAY, 50)
trigger.model.initiate()
Set the instrument to measure current.
Load the LoopUntilEvent trigger model to make measurements until the front panel trigger key is pressed,
then continue to make measurements equal to 50 % of the reading buffer size.
Also see
None