EasyManuals Logo
Home>Keithley>Multimeter>DMM7510

Keithley DMM7510 Reference Manual

Keithley DMM7510
1036 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #837 background imageLoading...
Page #837 background image
Section
8: TSP command reference Model DMM7510 7½ Digit Graphical Sampling Multimeter
Reference Manual
8-224 DMM7510-901-01 Rev. B / May 2015
Details
When the prompting mode is enabled, the instrument generates prompts when the instrument is
ready to take another command. Because the prompt is not generated until the previous command
completes, enabling prompts provides handshaking with the instrument to prevent buffer overruns.
When prompting is enabled, the instrument might generate the following prompts:
• TSP>. The standard prompt, which indicates that the previous command completed normally.
• TSP?. The prompt that is issued if there are unread entries in the event log when the prompt is
issued. Like the TSP> prompt, it indicates that processing of the command is complete. It does
not mean the previous command generated an error, only that there were still errors in the event
log when command processing completed.
• >>>>. The continuation prompt, which occurs when downloading scripts. When downloading
scripts, many command messages must be sent as a group. The continuation prompt indicates
that the instrument is expecting more messages as part of the present command.
Commands do not generate prompts. The instrument generates prompts in response to command
completion.
Prompts are enabled or disabled only for the remote interface that is active when you send the
command. For example, if you enable prompts when the GPIB connection is active, they will not be
enabled for a subsequent USB connection.
Do not disable prompting when using Test Script Builder. Test Script Builder requires prompts and
sets the prompting mode automatically. If you disable prompting, the instrument will stop responding
when you communicate using Test Script Builder because it is waiting for a common complete
prompt from Test Script Builder.
Example
localnode.prompts = localnode.ENABLE
Enable prompting.
Also see
tsplink.initialize() (on page 8-347)
localnode.prompts4882
This attribute enables and disables the generation of prompts for IEEE Std 488.2 common commands.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW)
Yes
Power cycle
Not saved
localnode.ENABLE
Usage
prompting = localnode.prompts4882
localnode.prompts4882 = prompting
prompting
IEEE Std 488.2 prompting mode:
• Disable prompting: localnode.DISABLE
•
Enable prompting: localnode.ENABLE

Table of Contents

Other manuals for Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM7510 and is the answer not in the manual?

Keithley DMM7510 Specifications

General IconGeneral
BrandKeithley
ModelDMM7510
CategoryMultimeter
LanguageEnglish

Related product manuals