6: SCPI command reference Model DMM7510 7½ Digit Graphical Sampling Multimeter
6-84 DMM7510-901-01 Rev. B / May 2015
[:SENSe[1]]:<function>:LINE:SYNC
This command determines if line synchronization is used during the measurement.
Type Affected by Where saved Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
0 (OFF)
Usage
[:SENSe[1]]:<function>:LINE:SYNC <state>
[:SENSe[1]]:<function>:LINE:SYNC?
The function to which the setting applies; see Functions
Disable:
or
Functions
Details
When line synchronization is enabled, measurements are initiated at the first positive-going zero
crossing of the power line cycle after the trigger.
Example
Turn on line synchronization when DC current is
measured.
Also see
Line cycle synchronization (on page 4-1)