6: SCPI command reference Model DMM7510 7½ Digit Graphical Sampling Multimeter
6-212 DMM7510-901-01 Rev. B / May 2015
:TRIGger:EXTernal:OUT:STIMulus
This command selects the event that causes a trigger to be asserted on the external output line.
Type Affected by Where saved Default value
Instrument reset
Usage
:TRIGger:EXTernal:OUT:STIMulus <event>
:TRIGger:EXTernal:OUT:STIMulus?
The event to use as a stimulus; see Details
Details
The trigger stimulus for the external output line can be set to one of the trigger events described in
the following table.
Trigger events
Event description Event constant
Front-panel TRIGGER key press
Notify trigger block <n> (1 to 8); the trigger model generates a
trigger event when it executes the notify block
A command interface trigger (bus trigger):
• Any remote interface: *TRG
• GPIB only: GET bus command
• VXI-11: VXI-11 command
Line edge (either rising, falling, or either based on the
configuration of the line) detected on digital input line <n> (1 to 6)
Line edge detected on TSP-Link synchronization line
(1 to 3)
Appropriate LXI trigger packet is received on LAN trigger object
(1 to 8)
Trigger event blender <n> (1 or 2), which combines trigger events
Trigger timer <n> (1 to 4) expired
Example
:TRIG:EXT:OUT:STIMulus TIM3
Set the stimulus for the external output to be
the expiration of trigger timer 3.
Also see