6: SCPI command reference Model DMM7510 7½ Digit Graphical Sampling Multimeter
6-108 DMM7510-901-01 Rev. B / May 2015
[:SENSe[1]]:<function>:THERmistor
This command describes the type of thermistor.
Type Affected by Where saved Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
5000
Usage
[:SENSe[1]]:<function>:THERmistor <n>
[:SENSe[1]]:<function>:THERmistor DEFault
[:SENSe[1]]:<function>:THERmistor MINimum
[:SENSe[1]]:<function>:THERmistor MAXimum
[:SENSe[1]]:<function>:THERmistor?
[:SENSe[1]]:<function>:THERmistor? DEFault
[:SENSe[1]]:<function>:THERmistor? MINimum
[:SENSe[1]]:<function>:THERmistor? MAXimum
The function to which the setting applies; see Functions
The thermistor type in ohms:
• 2252: 2252
• 5000: 5000
• 10000:
Functions
Details
This command is only applicable when the transducer type is set to thermistor.
For the <n> parameter, only 2252, 5000, or 10000 are valid entries. Other values you enter will cause
an out of range error message. The only exception to this is if you enter 2200 or 2250, the Model
DMM7510 will accept the entry but change it to 2252.
Example
TEMP:TRAN THER
TEMP:THER 2252
Set measurement function to temperature.
Set the transducer type to thermistor.
Set the thermistor type to 2252.
Also see
[:SENSe[1]]:<function>:TRANsducer (on page 6-112)
Temperature measurements (on page 2-120)