6: SCPI command reference Model DMM7510 7½ Digit Graphical Sampling Multimeter
6-172 DMM7510-901-01 Rev. B / May 2015
:TRACe:TRIGger:DIGitize
This command makes readings using the active digitize function and stores them in the reading buffer.
Type Affected by Where saved Default value
Command only Not applicable Not applicable Not applicable
Usage
:TRACe:TRIGger:DIGitize
:TRACe:TRIGger:DIGitize "<bufferName>"
A string that indicates the reading buffer; the default buffers
(defbuffer1 or defbuffer2) or the name of a user-
buffer is specified,
is used
Details
A digitize function must be selected before sending this command.
This command makes the number of digitize measurements that is set by the digitize count
command.
Example
TRACe:MAKE "MyBuffer", 100
TRACe:TRIG:DIG "MyBuffer"
TRACe:TRIG:DIG "MyBuffer"
TRACe:TRIG:DIG "MyBuffer"
TRACe:TRIG:DIG "MyBuffer"
TRACe:TRIG:DIG "MyBuffer"
TRACe:DATA? 1,5, "MyBuffer", rel
Make the digitize voltage measurement function the active
function.
Create a buffer called MyBuffer with a maximum size of 100.
Make readings and store them in MyBuffer.
Recall the relative time when the data points were measured
for the first five readings in the buffer.
Example output:
0.000000,0.408402,0.816757,1.208823,1.617529
Also see
[:SENSe[1]]:DIGitize:COUNt (on page 6-122)
[:SENSe[1]]:DIGitize:FUNCtion[:ON] (on page 6-123)
:TRACe:MAKE (on page 6-160)