Section
8: TSP command reference Model DMM7510 7½ Digit Graphical Sampling Multimeter
Reference Manual
8-346 DMM7510-901-01 Rev. B / May 2015
tsplink.group
This attribute contains the group number of a TSP-Link node.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW) Yes Not applicable Nonvolatile memory 0
Usage
groupNumber = tsplink.group
tsplink.group = groupNumber
groupNumber
The group number of the TSP-Link node (0 to 64)
Details
To remove the node from all groups, set the attribute value to 0.
When the node is turned off, the group number for that node changes to 0.
The master node can be assigned to any group. You can also include other nodes in the group that
includes the master. Note that any nodes that are set to 0 are automatically included in the group
that contains the master node, regardless of the group that is assigned to the master node.
Example
tsplink.group = 3
Assign the instrument to TSP-Link group number 3.
Also see
Using groups to manage nodes on a TSP-Link system (on page 3-110)

Table of Contents

Other manuals for Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM7510 and is the answer not in the manual?

Keithley DMM7510 Specifications

General IconGeneral
BrandKeithley
ModelDMM7510
CategoryMultimeter
LanguageEnglish

Related product manuals