7½ Digit Graphical Sampling Multimeter Reference Manual Section 8:
DMM7510-901-01 Rev. B / May 2015 8-159
dmm.measure.limit[Y].autoclear
This attribute indicates if the test result for limit Y should be cleared automatically or not.
Type TSP-Link accessible Affected by Where saved Default value
Instrument reset
Power cycle
Measure configuration list
Measure configuration list
Usage
value = dmm.measure.limit[Y].autoclear
dmm.measure.limit[Y].autoclear = value
The auto clear setting:
• Disable: dmm.OFF
• Enable:
Functions
dmm.FUNC_DIGITIZE_CURRENT
dmm.FUNC_DIGITIZE_VOLTAGE
Details
When auto clear is set to on for a measure function, limit conditions are cleared automatically after
each measurement. If you are making a series of measurements, the instrument shows the limit test
result of the last measurement for the pass or fail indication for the limit.
If you want to know if any of a series of measurements failed the limit, set the auto clear setting to off.
When this set to off, a failed indication is not cleared automatically. It remains set until it is cleared
with the clear command.
The auto clear setting affects both the high and low limits.
Example
See dmm.measure.limit[Y].low.value (on page 8-164) for an example of how to use this command.
Also see
dmm.digitize.limit[Y].autoclear (on page 8-93)
dmm.measure.limit[Y].enable (on page 8-161)