7½ Digit Graphical Sampling Multimeter Reference Manual Section 8:
DMM7510-901-01 Rev. B / May 2015 8-163
dmm.measure.limit[Y].high.value
This attribute specifies the upper limit for a limit test.
Type TSP-Link accessible Affected by Where saved Default value
Instrument reset
Power cycle
Measure configuration list
Measure configuration list
1 for most functions; see Details
Usage
highLimit = dmm.measure.limit[Y].high.value
dmm.measure.limit[Y].high.value = highLimit
The value of the upper limit (-1e+12 to 1e+12)
Functions
dmm.FUNC_DIGITIZE_CURRENT
dmm.FUNC_DIGITIZE_VOLTAGE
Details
This command sets the high limit for the limit Y test for the selected measurement function. When limit
Y testing is enabled, the instrument generates a fail indication when the measurement value is more
than this value.
Default is 0.8 for limit 1 when the diode function is selected; 10 when the continuity function is
selected. The default for limit 2 for the diode and continuity functions is 1.
Example
See dmm.measure.limit[Y].low.value (on page 8-164) for an example of how to use this command.
Also see
dmm.digitize.limit[Y].high.value (on page 8-99)
dmm.measure.limit[Y].enable (on page 8-161)
dmm.measure.limit[Y].low.value (on page 8-164)