7½ Digit Graphical Sampling Multimeter Reference Manual Section 8:
DMM7510-901-01 Rev. B / May 2015 8-193
dmm.measure.threertd
This attribute defines the type of three-wire RTD that is being used.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW) Yes
Restore configuration
Instrument reset
Power cycle
Measure configuration list
Configuration script
Measure configuration list
PT100
Usage
value = dmm.measure.threertd
dmm.measure.threertd = value
• PT100: dmm.RTD_PT100
• PT385: dmm.RTD_PT385
• PT3916: dmm.RTD_PT3916
• D100: dmm.RTD_D100
• F100: dmm.RTD_F100
•
User-specified type: dmm.RTD_USER
Functions
dmm.FUNC_DIGITIZE_CURRENT
dmm.FUNC_DIGITIZE_VOLTAGE
Details
The transducer type must be set to temperature and the transducer must be set to 3-wire RTD before
you can set the RTD type.
Example
dmm.measure.func = dmm.FUNC_TEMPERATURE
dmm.measure.transducer = dmm.TRANS_THREERTD
dmm.measure.threertd = dmm.RTD_D100
Set the measure function to temperature.
Set the transducer type to 3-wire RTD.
Set the RTD type to D100.
Also see
dmm.measure.rtdalpha (on page 8-184)
dmm.measure.rtdbeta (on page 8-185)
dmm.measure.rtddelta (on page 8-186)
dmm.measure.rtdzero (on page 8-187)
dmm.measure.transducer (on page 8-197)