:STATus:PRESet
This command resets all bits in the status model.
Type Affected by Where saved Default value
Command only
Not applicable
Not applicable
Not applicable
Usage
:STATus:PRESet
Details
This function clears the event registers and the enable registers for operation and questionable. It will
not clear the Service Request Enable Register (*SRE) to Standard Request Enable Register (*ESE).
Preset does not affect the event queue.
The Standard Event Status Register is not affected by this command.
Example
STAT:PRES
Resets the registers.
Also see
Status model (on page 1)
:STATus:QUEStionable:CONDition?
This command reads the Questionable Condition Register of the status model.
Type Affected by Where saved Default value
Query only
Not applicable
Not applicable
Not applicable
Usage
:STATus:QUEStionable:CONDition?
Details
This command reads the contents of the Questionable Condition Register, which is one of the
Questionable Event Registers.
For detail on interpreting the value of a register, see Understanding bit settings (on page 14
).
Example
:STAT:QUES:COND?
Reads the Questionable Condition Register.
Also see
Questionable Event Register (on page 7)
Understanding bit settings (on page 14)

Table of Contents

Other manuals for Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM7510 and is the answer not in the manual?

Keithley DMM7510 Specifications

General IconGeneral
BrandKeithley
ModelDMM7510
CategoryMultimeter
LanguageEnglish

Related product manuals