Setting autozero
You can enable or disable automatic referencing, or request a one-time refresh of the reference
values.
The reference setting is stored with the selected measure function.
To set autozero using the front panel:
1. Press the FUNCTION key.
2. Select the measure function.
3. Press the MENU key.
4. Under Measure, select Settings.
5. For Auto Zero, select On or Off.
6. If Off is selected, you can select the Zero Once option to send a one-time refresh.
7. Select HOME to return to the operating display.
To set autozero using SCPI commands:
Refer to the following commands:
• [:SENSe[1]]:<function>:AZERo[:STATe] (on page 6-72)
• [:SENSe[1]]:AZERo:ONCE (on page 6-114)
To set autozero using TSP commands:
Refer to the following commands:
• dmm.measure.autozero.enable (on page 8-135)
• dmm.measure.autozero.once() (on page 8-136)
Saving setups
You can save the present settings and any configuration lists that you have defined for the Model
DMM7510 to internal memory or an external USB flash drive.
After the settings are saved, you can recall the settings. You can also set them to be the default
settings when the instrument is powered on.
If you are using TSP commands, saved setups are scripts and can be added, modified, and deleted
like any other script. See Introduction to TSP operation (on page 7-1
) for additional information about
working with scripts.