7½ Digit Graphical Sampling Multimeter Reference Manual Section 6:
DMM7510-901-01 Rev. B / May 2015 6-81
[:SENSe[1]]:<function>:DELay:USER<n>
This command sets a user-defined delay that you can use in the trigger model.
Type Affected by Where saved Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
0
Usage
[:SENSe[1]]:<function>:DELay:USER<n> <delayTime>
[:SENSe[1]]:<function>:DELay:USER<n> DEFault
[:SENSe[1]]:<function>:DELay:USER<n> MINimum
[:SENSe[1]]:<function>:DELay:USER<n> MAXimum
[:SENSe[1]]:<function>:DELay:USER<n>?
[:SENSe[1]]:<function>:DELay:USER<n>? DEFault
[:SENSe[1]]:<function>:DELay:USER<n>? MINimum
[:SENSe[1]]:<function>:DELay:USER<n>? MAXimum
The function to which the setting applies; see Functions
The user delay to which this time applies (1 to 5)
The delay (0 for no delay, or 167 ns to 10 ks)
Functions
Details
To use this command in a trigger model, assign the delay to the dynamic delay block.
The delay is specific to the selected function.
Example
:TRIGger:BLOCk:DELay:DYNamic 6, MEAS1
Set user delay 1 to 0.2 s for current
measurements. Set trigger block 6 to be a
dynamic delay that is set to user delay 1 for the
Also see
:TRIGger:BLOCk:DELay:DYNamic (on page 6-195)