7½ Digit Graphical Sampling Multimeter Reference Manual Section 6:
SCPI command reference
DMM7510-901-01 Rev. B / May 2015 6-15
Example
*SAV 2
Saves the instrument settings in mem
ory location
2.
Also see
Saving setups (on page 2-150)
*RCL (on page 6-14)
ACAL subsystem
Automatic calibration removes measurement errors that are caused by the performance drift on the
components used in the DMM as a result of temperature and time.
:ACAL:COUNt?
This command returns the number of times automatic calibration has been run.
Type Affected by Where saved Default value
Query only
Not applicable
Nonvolatile memory
Not applicable
Usage
:ACAL:COUNt?
Details
The number of times that auto calibration has been run since the last factory calibration. The count
restarts at 1 after a factory calibration.
Example
ACAL:COUN?
Returns the number of times auto calibration has
been run.
Example output:
15
Also see
Auto calibration (on page 3-44)
:ACAL:RUN (on page 6-20)

Table of Contents

Other manuals for Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM7510 and is the answer not in the manual?

Keithley DMM7510 Specifications

General IconGeneral
BrandKeithley
ModelDMM7510
CategoryMultimeter
LanguageEnglish

Related product manuals