7½ Digit Graphical Sampling Multimeter Reference Manual Section 8:
DMM7510-901-01 Rev. B / May 2015 8-267
Trigger events
Event description Event constant
Trigger event blender N (1 to 2), which combines trigger
events
Trigger timer N (1 to 4) expired
trigger.EVENT_ANALOGTRIGGER
External in trigger
Example
digio.line[2].mode = digio.MODE_TRIGGER_OUT
trigger.digout[2].stimulus = trigger.EVENT_TIMER3
Set the stimulus for output digital
trigger line 2 to be the expiration of
Also see
digio.line[N].mode (on page 8-52)
digio.line[N].reset() (on page 8-54)
Digital I/O port configuration (on page 3-49)
trigger.digin[N].clear() (on page 8-260)
trigger.digout[N].assert() (on page 8-263)
trigger.ext.reset()
This function resets the edge, logic, and stimulus values for the external in/out port to their default values.
Type TSP-Link accessible
Affected by Where saved Default value
Usage
trigger.ext.reset()
Details
This function resets the following attributes to their default values:
• trigger.extin.edge
• trigger.extout.logic
• trigger.extout.stimulus
It also clears trigger.extin.overrun.