EasyManuals Logo
Home>Keithley>Multimeter>DMM7510

Keithley DMM7510 Reference Manual

Keithley DMM7510
1036 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #880 background imageLoading...
Page #880 background image
Model DMM7510
7½ Digit Graphical Sampling Multimeter Reference Manual Section 8:
TSP command reference
DMM7510-901-01 Rev. B / May 2015 8-267
Trigger events
Event description Event constant
No trigger event
trigger.EVENT_NONE
Trigger event blender N (1 to 2), which combines trigger
events
trigger.EVENT_BLENDERN
Trigger timer N (1 to 4) expired
trigger.EVENT_TIMERN
Analog trigger
trigger.EVENT_ANALOGTRIGGER
External in trigger
trigger.EVENT_EXTERNAL
Example
digio.line[2].mode = digio.MODE_TRIGGER_OUT
trigger.digout[2].stimulus = trigger.EVENT_TIMER3
Set the stimulus for output digital
trigger line 2 to be the expiration of
trigger timer 3.
Also see
digio.line[N].mode (on page 8-52)
digio.line[N].reset() (on page 8-54)
Digital I/O port configuration (on page 3-49)
trigger.digin[N].clear() (on page 8-260)
trigger.digout[N].assert() (on page 8-263)
trigger.ext.reset()
This function resets the edge, logic, and stimulus values for the external in/out port to their default values.
Type TSP-Link accessible
Affected by Where saved Default value
Function
Yes
Usage
trigger.ext.reset()
Details
This function resets the following attributes to their default values:
• trigger.extin.edge
• trigger.extout.logic
• trigger.extout.stimulus
It also clears trigger.extin.overrun.

Table of Contents

Other manuals for Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM7510 and is the answer not in the manual?

Keithley DMM7510 Specifications

General IconGeneral
BrandKeithley
ModelDMM7510
CategoryMultimeter
LanguageEnglish

Related product manuals