Model DMM7510
7½ Digit Graphical Sampling Multimeter Reference Manual Section 6: SCPI command referen
ce
DMM7510-901-01 Rev. B / May 2015 6-117
Details
Deletes a configuration list. If the index is not specified, the entire configuration list is deleted. If the
index is specified, only the specified configuration index in the list is deleted.
When an index is deleted from a configuration list, the index numbers of the following indexes are
shifted up by one. For example, if you have a configuration list with 10 indexes and you delete index
3, the index that was numbered 4 becomes index 3, and the all the following indexes are renumbered
in sequence to index 9. Because of this, if you want to delete several nonconsecutive indexes in a
configuration list, it is best to delete the higher numbered index first, then the next lower index, and so
on. This also means that if you want to delete all the indexes in a configuration list, you must delete
index 1 repeatedly until all indexes have been removed.
Example
:SENSe:CONF:LIST:DELete "myMeasList"
Deletes a configuration list named myMeasList.
:SENSe:CONF:LIST:DELete "myMeasList", 2
Deletes configuration index 2 in a configuration list
named myMeasList.
Also see
Configuration lists (on page 3-37)
[:SENSe[1]]:CONFiguration:LIST:CREate (on page 6-116)
[:SENSe[1]]:CONFiguration:LIST:QUERy?
This command returns a list of TSP commands and parameter settings that are stored in the specified
configuration index.
Type Affected by Where saved Default value
Query only
Not applicable
Not applicable
Not applicable
Usage
[:SENSe[1]]:CONFiguration:LIST:QUERy? "<name>", <index>
[:SENSe[1]]:CONFiguration:LIST:QUERy? "<name>", <index>, <fieldSeparator>
<name>
A string that represents the name of a measure configuration list
<index>
A number that defines a specific configuration index in the configuration list
<fieldSeparator>
A separator for the data:
• Comma (default): 1
• Semicolon: 2
• New line:
3

Table of Contents

Other manuals for Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM7510 and is the answer not in the manual?

Keithley DMM7510 Specifications

General IconGeneral
BrandKeithley
ModelDMM7510
CategoryMultimeter
LanguageEnglish

Related product manuals