Model DMM7510
7½ Digit Graphical Sampling Multimeter Reference Manual Section 8:
TSP command reference
DMM7510-901-01 Rev. B / May 2015 8-189
dmm.measure.sense.range
This attribute determines the positive full-scale range for the sense measurement.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW)
Yes
Restore configuration
Instrument reset
Power cycle
Measure configuration list
Configuration script
Measure configuration list
10
Usage
value = dmm.measure.sense.range
dmm.measure.sense.range = value
value
Range in volts:
0.1
,
1
, or
10
dmm.FUNC_DC_VOLTAGE
dmm.FUNC_RESISTANCE
dmm.FUNC_ACV_FREQUENCY
dmm.FUNC_AC_VOLTAGE
dmm.FUNC_4W_RESISTANCE
dmm.FUNC_ACV_PERIOD
dmm.FUNC_DC_CURRENT
dmm.FUNC_DIODE
dmm.FUNC_DCV_RATIO
dmm.FUNC_AC_CURRENT
dmm.FUNC_CAPACITANCE
dmm.FUNC_DIGITIZE_CURRENT
dmm.FUNC_TEMPERATURE
dmm.FUNC_CONTINUITY
dmm.FUNC_DIGITIZE_VOLTAGE
Details
Determines the full-scale input for the reference measurement in the denominator of the ratio. It also
affects the accuracy of the measurements and the maximum signal that can be measured. Autorange
is automatically set to off if a specific value is set.
When you assign a range value, the instrument is set on a fixed range that is large enough to
measure the assigned value. The instrument selects the best range for measuring the maximum
expected value.
For example, if you expect a sense reading of approximately 9 V, set the range to 9 V to select the
10 V range.
This command is primarily intended to eliminate the time that is required by the instrument to select
an automatic range.
Note that when you select a fixed range, an overflow condition can occur.
When you read this setting, you see the positive full-scale value of the sense range that the
instrument is presently using.
Example
dmm.measure.func = dmm.FUNC_DCV_RATIO
dmm.measure.sense.range = 1
Select the DC voltage ratio function.
Set the sense range to 1 V.
Also see
Ranges (on page 3-3)
dmm.measure.sense.autorange (on page 8-188)

Table of Contents

Other manuals for Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM7510 and is the answer not in the manual?

Keithley DMM7510 Specifications

General IconGeneral
BrandKeithley
ModelDMM7510
CategoryMultimeter
LanguageEnglish

Related product manuals