Model DMM7510
7½ Digit Graphical Sampling Multimeter Reference Manual Section 6:
SCPI command reference
DMM7510-901-01 Rev. B / May 2015 6-155
:TRACe:DATA?
This command returns specified data elements from a specified reading buffer.
Type Affected by Where saved Default value
Query only
Recall settings
Instrument reset
Power cycle
Not applicable
Not applicable
Usage
:TRACe:DATA? <startIndex>, <endIndex>
:TRACe:DATA? <startIndex>, <endIndex>, "<bufferName>"
:TRACe:DATA? <startIndex>, <endIndex>, "<bufferName>", <bufferElements>
:TRACe:DATA? <startIndex>, <endIndex>, "<bufferName>", <bufferElements>,
<bufferElements>
<startIndex>
Beginning index of the buffer to return; must be 1 or greater
<endIndex>
Ending index of the buffer to return
<bufferName>
A string that indicates the reading buffer; the default buffers (defbuffer1 or
defbuffer2) or the name of a user-defined buffer; if no buffer is specified,
defbuffer1
is used
<bufferElements>
A list of elements in the buffer to print; if nothing is specified, READing is used; see
Details for the list of options for buffer elements; a maximum of 14 comma-delimited
buffer elements may be specified
Details
The output of :TRACe:DATA? is affected by the data format selected by :FORMat[:DATA]. If you
set FORMat[:DATA] to REAL or SREAL, you will have fewer options for buffer elements. The only
buffer elements available are READing, RELative, and EXTRa. If you request a buffer element that
is not permitted for the selected data format, the instrument generates the error 1133, "Parameter 4,
Syntax error, expected valid name parameters."
To change the number of digits returned in a remote command reading, use the
:FORMat:ASCii:PRECision command.
When specifying buffer elements, you can:
• Specify buffer elements in any order.
• Include up to 12 elements in a single list. You can repeat elements as long as the number of
elements in the list is less than 12.
• Use a comma to delineate multiple elements for a data point.

Table of Contents

Other manuals for Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM7510 and is the answer not in the manual?

Keithley DMM7510 Specifications

General IconGeneral
BrandKeithley
ModelDMM7510
CategoryMultimeter
LanguageEnglish

Related product manuals