6: SCPI command reference Model DMM7510 7½ Digit Graphical Sampling Multimeter
6-208 DMM7510-901-01 Rev. B / May 2015
:TRIGger:DIGital<n>:OUT:STIMulus
This command selects the event that causes a trigger to be asserted on the digital output line.
Type Affected by Where saved Default value
Instrument reset
Usage
:TRIGger:DIGital<n>:OUT:STIMulus <event>
:TRIGger:DIGital<n>:OUT:STIMulus?
Digital I/O trigger line (1 to 6)
The event to use as a stimulus; see Details
Details
The digital trigger pulsewidth command determines how long the trigger is asserted. The trigger
stimulus for a digital I/O line can be set to one of the trigger events that are described in the following
table.
Trigger events
Event description Event constant
Front-panel TRIGGER key press
Notify trigger block <n> (1 to 8); the trigger model generates a trigger event when it
executes the notify block
A command interface trigger (bus trigger):
• Any remote interface: *TRG
• GPIB only: GET bus command
• VXI-11: VXI-11 command
Line edge (either rising, falling, or either based on the configuration of the line)
detected on digital input line
(1 to 6)
Line edge detected on TSP-Link synchronization line <n> (1 to 3)
Appropriate LXI trigger packet is received on LAN trigger object <n> (1 to 8)
Trigger event blender <n> (1 or 2), which combines trigger events
Trigger timer
(1 to 4) expired
Analog trigger
Example
:TRIG:DIG2:OUT:STIMulus TIM3
Set the stimulus for output digital trigger line
2 to be the expiration of trigger timer 3.
Also see
Digital I/O port configuration (on page 3-49)
:DIGital:LINE<n>:STATe (on page 6-40)
:TRIGger:DIGital<n>:OUT:LOGic (on page 6-206)