EasyManuals Logo
Home>Keithley>Multimeter>DMM7510

Keithley DMM7510 Reference Manual

Keithley DMM7510
1036 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #535 background imageLoading...
Page #535 background image
Section
6: SCPI command reference Model DMM7510 7½ Digit Graphical Sampling Multimeter
Reference Manual
6-208 DMM7510-901-01 Rev. B / May 2015
:TRIGger:DIGital<n>:OUT:STIMulus
This command selects the event that causes a trigger to be asserted on the digital output line.
Type Affected by Where saved Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
NONE
Usage
:TRIGger:DIGital<n>:OUT:STIMulus <event>
:TRIGger:DIGital<n>:OUT:STIMulus?
<n>
Digital I/O trigger line (1 to 6)
<event>
The event to use as a stimulus; see Details
Details
The digital trigger pulsewidth command determines how long the trigger is asserted. The trigger
stimulus for a digital I/O line can be set to one of the trigger events that are described in the following
table.
Trigger events
Event description Event constant
No trigger event
NONE
Front-panel TRIGGER key press
DISPlay
Notify trigger block <n> (1 to 8); the trigger model generates a trigger event when it
executes the notify block
NOTify<n>
A command interface trigger (bus trigger):
• Any remote interface: *TRG
• GPIB only: GET bus command
• VXI-11: VXI-11 command
device_trigger
COMMand
Line edge (either rising, falling, or either based on the configuration of the line)
detected on digital input line
<n>
(1 to 6)
DIGio<n>
Line edge detected on TSP-Link synchronization line <n> (1 to 3)
TSPLink<n>
Appropriate LXI trigger packet is received on LAN trigger object <n> (1 to 8)
LAN<n>
Trigger event blender <n> (1 or 2), which combines trigger events
BLENder<n>
Trigger timer
<n>
(1 to 4) expired
TIMer<n>
Analog trigger
ATRigger
External in trigger
EXTernal
Example
:TRIG:DIG2:OUT:STIMulus TIM3
Set the stimulus for output digital trigger line
2 to be the expiration of trigger timer 3.
Also see
Digital I/O port configuration (on page 3-49)
:DIGital:LINE<n>:STATe (on page 6-40)
:TRIGger:DIGital<n>:OUT:LOGic (on page 6-206)

Table of Contents

Other manuals for Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM7510 and is the answer not in the manual?

Keithley DMM7510 Specifications

General IconGeneral
BrandKeithley
ModelDMM7510
CategoryMultimeter
LanguageEnglish

Related product manuals