7½ Digit Graphical Sampling Multimeter Reference Manual Section 8:
DMM7510-901-01 Rev. B / May 2015 8-291
The bit pattern that is sent when the measurement fails limit 1; range 1 to
The bit pattern that is sent when the measurement fails limit 2; range 1 to
The bit pattern that is sent when the measurement fails limit 3; range 1 to
15; default is 4
The bit pattern that is sent when the measurement fails limit 4; range 1 to
15; default is 8
The bit pattern that is sent when all limits have passed; 1 to 15; default
is 15
The name of the reading buffer, which may be a default buffer
(defbuffer1 or defbuffer2) or a user-defined buffer; defaults to
Details
This trigger model template allows you to grade components and place them into up to four bins,
based on the comparison to limits.
To set a limit as unused, set the high value for the limit to be less than the low limit.
All limit patterns and the pass pattern are sent on digital I/O lines 1 to 4, where 1 is the least
significant bit.
The out line of the EXT TRIG IN/OUT rear-panel terminal is asserted at the end of each
measurement.
After selecting a trigger model template, you can view the trigger model blocks in a graphical format
by pressing the front-panel MENU key and under Trigger, selecting Configure. You can also add or
delete blocks and change trigger model settings from this screen. You can use the
trigger.model.getblocklist() command to view the trigger model blocks in a list format.
Example
For a detailed example, see the section in the Model DMM7510 User's Manual named "Grading and
binning resistors."
Also see
None