4: Source-measure considerations  Model 2461 Interactive SourceMeterĀ® Instrument 
 
4-24  2461-901-01 A/November 2015 
 
Moving average filter 
When the moving average filter is selected, the measurements are added to the stack continuously 
on a first-in, first-out basis. As each measurement is made, the oldest measurement is removed from 
the stack. A new averaged sample is produced using the new measurement and the data that is now 
in the stack. 
Note that when the moving average filter is first selected, the stack is empty. When the first 
measurement is made, it is copied into all the stack locations to fill the stack. A true average is not 
produced until the stack is filled with new measurements. 
For example, if the filter size is four, the first measurement is copied to all four stack locations. 
Therefore, (Reading1 + Reading1 + Reading1 + Reading1)/4. The display and remote interface 
update after first reading. With each additional measurement, the average updates: 
(Reading2 + Reading1 + Reading1 + Reading1)/4 
(Reading3 + Reading2 + Reading1 + Reading1)/4 
(Reading4 + Reading3 + Reading2 + Reading1)/4 
Do not use the moving average filter when performing a sweep in which source levels are being 
changed. You should always use a repeating average filter in this case. 
 
Setting up the averaging filter 
Using the front panel: 
1.  Press the MENU key. 
2.  Under Measure, select Calculations. 
3.  For Filter, select On to enable filtering. 
4.  Select Config. 
5.  For the Filter Type, select Moving or Repeat. 
6.  For the Filter Count, enter the number of measurements to be made for each averaged 
measurement sample. 
7.  Select OK. 
8.  Select HOME to return to the Home screen to view the measurements with the filter applied. 
Once the filter is set up, you can enable and disable the filter from the SETTINGS swipe screen. 
When filtering is enabled, the FILT indicator on the Home screen is lit. 
 
Using SCPI commands: 
To set the averaging filters using SCPI commands, refer to the following command descriptions: 
[:SENSe[1]]:<function>:AVERage:COUNt (on page 6-53) 
[:SENSe[1]]:<function>:AVERage[:STATe] (on page 6-55) 
[:SENSe[1]]:<function>:AVERage:TCONtrol (on page 6-56) 
 
Using TSP commands: 
To set the averaging filters using TSP commands, refer to the following command descriptions: