Model 2461
Interactive SourceMeter® Instrument Reference Manual Section 8:
TSP command reference
2461-901-01 A/November 2015 8-11
Example
buffer.clearstats()
Clears statistics for
defbuffer1
.
buffer.clearstats(testData)
Clears statistics for testData.
Also see
buffer.getstats() (on page 8-12)
buffer.delete()
This function deletes a user-defined reading buffer.
Type TSP-Link accessible
Affected by Where saved Default value
Function Yes
Usage
buffer.delete(bufferName)
bufferName
The name of a user-defined reading buffer
Details
You cannot delete the default reading buffers, defbuffer1 and defbuffer2.
Example
buf400 = buffer.make(400)
smu.measure.read(buf400)
printbuffer(1, buf400.n, buf400.relativetimestamps)
buffer.delete(buf400)
Create a 400-element reading buffer named
buf400
.
Make measurements and store the readings in buf400.
Print the relative timestamps for each reading in the buffer.
Example output, assuming five readings are stored in the buffer:
0, 0.412850017, 0.821640085, 1.230558058, 1.629523236
Delete buf400.
Also see
buffer.make() (on page 8-14)
bufferVar.clear() (on page 8-19)
printbuffer() (on page 8-93)
Reading buffers (on page 3-2)
Remote buffer operation (on page 3-23)

Table of Contents

Other manuals for Keithley 2461

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2461 and is the answer not in the manual?

Keithley 2461 Specifications

General IconGeneral
BrandKeithley
Model2461
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals