Section
6: SCPI command reference Model 2461 Interactive SourceMeter® Instrument
Reference Manual
6-222 2461-901-01 A/November 2015
:TRIGger:LOAD "ConfigList"
This command loads a predefined trigger model configuration that uses source and measure configuration lists.
Type Affected by Where saved Default value
Command only Not applicable Not applicable Not applicable
Usage
:TRIGger:LOAD "ConfigList", "<measureConfigList>", "<sourceConfigList>"
:TRIGger:LOAD "ConfigList", "<measureConfigList>", "<sourceConfigList>", <delay>
:TRIGger:LOAD "ConfigList", "<measureConfigList>", "<sourceConfigList>", <delay>,
"<bufferName>"
:TRIGger:LOAD "ConfigList", "<measureConfigList>", "<sourceConfigList>", <delay>,
"<bufferName>", <readingBlock>
<measureConfigList>
A string that contains the name of the measurement configuration list to
use
<sourceConfigList>
A string that contains the name of the source configuration list to use
<delay>
The delay time before each measurement (167 ns to 10 ks); default is 0 for
no delay
<bufferName>
A string that indicates the reading buffer; the default buffers (
defbuffer1
or defbuffer2) or the name of a user-defined buffer; if no buffer is
specified,
defbuffer1
is used
<readingBlock>
Define a measure or digitize block for the trigger model; options are:
ACTive: Add a measure or digitize block to the trigger model based
on the active function; if no option defined, ACTive is used
MEASure: Adds a measure block to the trigger model
DIGitize
: Adds a digitize block to the trigger model
Details
This trigger model template incorporates a source configuration list and measure configuration list.
You must set up the configuration lists before loading the trigger model.
You can also set a delay and change the reading buffer.
After selecting a trigger model template, you can view the trigger model blocks in a graphical format
by pressing the front-panel MENU key and under Trigger, selecting Configure. You can also add or
delete blocks and change trigger model settings from this screen. You can use the
TRIGger:BLOCk:LIST? command to view the trigger model blocks in a list format.
This command replaces the TRIGger:LOAD:CONFiguration:LIST command, which is
deprecated.

Table of Contents

Other manuals for Keithley 2461

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2461 and is the answer not in the manual?

Keithley 2461 Specifications

General IconGeneral
BrandKeithley
Model2461
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals