Section
8: TSP command reference Model 2461 Interactive SourceMeter® Instrument
Reference Manual
8-134 2461-901-01 A/November 2015
smu.measure.autozero.enable
This attribute enables or disables automatic updates to the internal reference measurements (autozero) of the
instrument.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW)
Yes
Restore configuration
Instrument reset
Power cycle
Measure configuration list
Function change
Configuration script
Measure configuration list
smu.ON
Usage
state = smu.measure.autozero.enable
smu.measure.autozero.enable = state
state
The status of autozero; set to one of the following values:
Disable autozero: smu.OFF
Enable autozero: smu.ON
Details
To ensure the accuracy of readings, the instrument must periodically get new measurements of its
internal ground and voltage reference. The time interval between updates to these reference
measurements is determined by the integration aperture that is being used for measurements. The
Model 2461 uses separate reference and zero measurements for each aperture.
By default, the instrument automatically checks these reference measurements whenever a signal
measurement is made.
The time to make the reference measurements is in addition to the normal measurement time. If
timing is critical, such as in sweeps, you can disable autozero to avoid this time penalty.
When autozero is set to off, the instrument may gradually drift out of specification. To minimize the
drift, you can send the once command to make a reference and zero measurement immediately
before a test sequence.
Example
smu.measure.func = smu.FUNC_DC_VOLTAGE
smu.measure.autozero.enable = smu.OFF
Set autozero off for voltage measurements.
Also see
smu.measure.autozero.once() (on page 8-135)
smu.measure.nplc (on page 8-164)

Table of Contents

Other manuals for Keithley 2461

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2461 and is the answer not in the manual?

Keithley 2461 Specifications

General IconGeneral
BrandKeithley
Model2461
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals