Section
6: SCPI command reference Model 2461 Interactive SourceMeter® Instrument
Reference Manual
6-156 2461-901-01 A/November 2015
Example
syst:time 2014, 8, 29, 11, 30, 30
syst:time? 1
Set the system time to August 29, 2014 at
11:30:30 and confirm setting.
Output:
Fri Aug 29 11:30:37 2014
Also see
None
:SYSTem:VERSion?
Query the present SCPI version.
Type Affected by Where saved Default value
Query only
Not applicable
Not applicable
Not applicable
Usage
:SYSTem:VERSion?
Details
This query command returns the SCPI version.
Example
SYSTem:VERSion?
Query the version. An example of a return is:
1996.0
Also see
None
TRACe subsystem
The TRACe subsystem contains commands that control the reading buffers.
:TRACe:ACTual?
This command contains the number of readings in the specified reading buffer.
Type Affected by Where saved Default value
Query only
Recall settings
Instrument reset
Power cycle
Not applicable
Not applicable
Usage
:TRACe:ACTual?
:TRACe:ACTual? "<bufferName>"
<bufferName>
A string that indicates the reading buffer; the default buffers (
defbuffer1
or
defbuffer2) or the name of a user-defined buffer; if no buffer is specified,
defbuffer1 is used

Table of Contents

Other manuals for Keithley 2461

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2461 and is the answer not in the manual?

Keithley 2461 Specifications

General IconGeneral
BrandKeithley
Model2461
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals