Interactive SourceMeter® Instrument Reference Manual Section 6:
2461-901-01 A/November 2015 6-45
:FORMat:BORDer
This command sets the byte order for the IEEE Std. 754 binary formats.
Type Affected by Where saved Default value
Command and query
Recall settings
Instrument reset
Save settings SWAP
Usage
:FORMat:BORDer <name>
:FORMat:BORDer?
• Normal byte order: NORMal
•
Reverse byte order for binary formats: SWAPped
Details
This attribute selected the byte order in which data is written.
The SWAPped byte order must be used when transmitting binary data to a computer with a Microsoft
Windows operating system.
The ASCII data format can only be sent in the normal byte order. If the ASCII format is selected, the
SWAPped selection is ignored.
When you select NORMal byte order, the data format for each element is sent as follows:
Byte 1 Byte 2 Byte 3 Byte 4
(Single precision)
When you select SWAPped, the data format for each element is sent as follows:
Byte 4 Byte 3 Byte 2 Byte 1
(Single precision)
The #0 header is not affected by this command. The header is always sent at the beginning of the
data string for each measurement conversion.
Example
Use the normal byte order.
Also see
:FORMat[:DATA] (on page 6-46)