6: SCPI command reference Model 2461 Interactive SourceMeter® Instrument
6-240 2461-901-01 A/November 2015
:TRIGger:TIMer<n>:STARt:OVERrun?
This command indicates if an event was ignored because of the event detector state.
Type Affected by Where saved Default value
Usage
:TRIGger:TIMer<n>:STARt:OVERrun?
Trigger timer number (1 to 4)
Details
This command indicates if an event was ignored because the event detector was already in the
detected state when the event occurred.
This is an indication of the state of the event detector built into the timer itself. It does not indicate if
an overrun occurred in any other part of the trigger model or in any other construct that is monitoring
the delay completion event. It also is not an indication of a delay overrun.
This returns 0 if there is no overrun or 1 if there is an overrun.
Example
Checks the overrun status on trigger timer 1.
Also see
None
:TRIGger:TIMer<n>:STARt:SEConds
This command configures an alarm or a time in the future when the timer will start.
Type Affected by Where saved Default value
Instrument reset
Power cycle
Usage
:TRIGger:TIMer<n>:STARt:SEConds <time>
:TRIGger:TIMer<n>:STARt:SEConds?
Trigger timer number (1 to 4)
The time: 0 to 2,147,483,647 s