Section
6: SCPI command reference Model 2461 Interactive SourceMeter® Instrument Reference Man
ual
6-104 2461-901-01 A/November 2015
:SOURce[1]:PULSe:<function>:<x>LIMit[:LEVel]
This command sets the source limit for pulsed output for the selected function when pulsing.
Type Affected by Where saved Default value
Command and query
Recall settings
Instrument reset
Power cycle
Source configuration list
Save settings
Source configuration list
Current: 105 µA
Voltage: 7.35 V
Usage
:SOURce[1]:PULSe:CURRent:VLIMit[:LEVel] <n>
:SOURce[1]:PULSe:CURRent:VLIMit[:LEVel]?
:SOURce[1]:PULSe:CURRent:VLIMit[:LEVel]? DEFault
:SOURce[1]:PULSe:CURRent:VLIMit[:LEVel]? MINimum
:SOURce[1]:PULSe:CURRent:VLIMit[:LEVel]? MAXimum
:SOURce[1]:PULSe:VOLTage:ILIMit[:LEVel] <n>
:SOURce[1]:PULSe:VOLTage:ILIMit[:LEVel]?
:SOURce[1]:PULSe:VOLTage:ILIMit[:LEVel]? DEFault
:SOURce[1]:PULSe:VOLTage:ILIMit[:LEVel]? MINimum
:SOURce[1]:PULSe:VOLTage:ILIMit[:LEVel]? MAXimum
<function>
The source function to which this setting applies:
Current: CURRent
Voltage:
VOLTage
<x>
The function to which the limit applies:
Current: I
Voltage:
V
<n>
The limit:
Current: 10 nA to 10.5 A
Voltage: 2 mV to 105 V
Details
This command sets the source limit for pulse outputs. The Model 2461 cannot source pulse levels
that exceed this limit.
The values that can be set for this command are limited by the setting for the overvoltage protection
limit. Also, if an interlock is not installed, the values are limited.
This value can also be limited by the measurement range. If a specific measurement range is set, the
limit must be more than 1 percent of the measurement range. If you set the measurement range to be
automatically selected, the measurement range is selected based on the pulse and bias limits.
If you change the source range to a level that is not appropriate for this limit, the instrument changes
the source limit to a limit that is appropriate to the range and a warning is generated.
Limits are absolute values.

Table of Contents

Other manuals for Keithley 2461

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2461 and is the answer not in the manual?

Keithley 2461 Specifications

General IconGeneral
BrandKeithley
Model2461
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals