Interactive SourceMeter® Instrument Reference Manual Section 8:
2461-901-01 A/November 2015 8-85
localnode.model
This attribute stores the model number.
Type TSP-Link accessible Affected by Where saved Default value
Usage
model = localnode.model
The model number of the instrument
Example
Outputs the model number of the local node. For example:
2461
Also see
localnode.serialno (on page 8-87)
localnode.password
This attribute stores the instrument password.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (W) No Rear-panel LAN reset Nonvolatile memory "admin"
Usage
localnode.password = "password"
A string that contains the instrument password (maximum 30 characters)
Details
When the access to the instrument is set to protected or lockout, this is the password that is used to
gain access.
If you forget the password, you can reset the password to the default:
1. On the front panel, press MENU.
2. Under System, select Info/Manage.
3. Select Password Reset.
You can also reset the password and the LAN settings from the rear panel by inserting a
straightened paper clip into hole below LAN RESET.
Example
localnode.password = "N3wpa55w0rd"
Changes the password to N3wpa55w0rd.
Also see
localnode.access (on page 8-83)