Section
6: SCPI command reference Model 2461 Interactive SourceMeter® Instrument
Reference Manual
6-148 2461-901-01 A/November 2015
:SYSTem:ERRor:CODE[:NEXT]?
This command reads the oldest error code.
Type Affected by Where saved Default value
Query only
Not applicable
Not applicable
Not applicable
Usage
:SYSTem:ERRor:CODE[:NEXT]?
Details
This command returns the numeric code of the next error in the event log. The error is cleared from
the queue after being read.
This command returns only error messages from the event log. To return information and warning
messages, see :SYSTem:EVENtlog:NEXT?
Example
SYST:ERR:CODE?
Returns the error code of the next error in
the event log.
For example, if error -222, Parameter data
out of range error, occurred, the output is:
-222
Also see
:SYSTem:EVENtlog:NEXT? (on page 6-150)
:SYSTem:ERRor:COUNt?
This command returns the number of errors in the event log.
Type Affected by Where saved Default value
Query only
Not applicable
Not applicable
Not applicable
Usage
:SYSTem:ERRor:COUNt?
Details
This command does not return other types of events, such as information messages. To return other
types of events, use :SYSTem:EVENtlog:COUNt?
This command does not clear the errors from the event log.
Example
SYST:ERR:COUN?
If there are five errors in the event log, the
output is:
5
Also see
:SYSTem:EVENtlog:COUNt? (on page 6-149)

Table of Contents

Other manuals for Keithley 2461

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2461 and is the answer not in the manual?

Keithley 2461 Specifications

General IconGeneral
BrandKeithley
Model2461
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals