6: SCPI command reference Model 2461 Interactive SourceMeter® Instrument
6-138 2461-901-01 A/November 2015
:STATus:QUEStionable[:EVENt]?
This command reads the Questionable Event Register.
Type Affected by Where saved Default value
Usage
:STATus:QUEStionable[:EVENt]?
Details
This query reads the contents of the questionable status event register. After sending this command
and addressing the instrument to talk, a value is sent to the computer. This value indicates which bits
in the appropriate register are set.
The Questionable Register can be set to the numeric equivalent of the bit to set. To set more than
one bit of the register, set the Questionable Register to the sum of their decimal weights. For
example, to set bits B12 and B13, set the Questionable Register to 12,288 (which is the sum of 4,096
+ 8,192).
Example
Query the Questionable Register.
Also see
Questionable Event Register (on page C-7)