6: SCPI command reference Model 2461 Interactive SourceMeter® Instrument
6-126 2461-901-01 A/November 2015
:SOURce[1]:SWEep:<function>:LINear:STEP
This command sets up a linear source sweep configuration list and trigger model with a fixed number of steps.
Type Affected by Where saved Default value
Command only
Recall settings
Instrument reset
Save settings Not applicable
Usage
:SOURce[1]:SWEep:<function>:LINear:STEP <start>, <stop>, <steps>
:SOURce[1]:SWEep:<function>:LINear:STEP <start>, <stop>, <steps>, <delay>
:SOURce[1]:SWEep:<function>:LINear:STEP <start>, <stop>, <steps>, <delay>, <count>
:SOURce[1]:SWEep:<function>:LINear:STEP <start>, <stop>, <steps>, <delay>, <count>,
<rangeType>
:SOURce[1]:SWEep:<function>:LINear:STEP <start>, <stop>, <steps>, <delay>, <count>,
<rangeType>, <failAbort>
:SOURce[1]:SWEep:<function>:LINear:STEP <start>, <stop>, <steps>, <delay>, <count>,
<rangeType>, <failAbort>, <dual>
:SOURce[1]:SWEep:<function>:LINear:STEP <start>, <stop>, <steps>, <delay>, <count>,
<rangeType>, <failAbort>, <dual>, "<bufferName>"
Current sweep:
The voltage or current source level at which the sweep starts:
• Current: -7.35 A to 7.35 A
•
The voltage or current at which the sweep stops:
• Current: -7.35 A to 7.35 A
•
The step size at which the source level will change; step size must be greater than 0
The delay between measurement points; default is -1, which enables autodelay, or
a specific delay value from 50
µ
s to 10,000 s; or 0 for no delay
The number of times to run the sweep; default is
:
• Infinite loop: 0
•
Finite loop: 1 to 268435455
The source range that is used for the sweep:
• Most sensitive source range for each source level in the sweep: AUTO
• Best fixed range: BEST (default)
• Present source range for the entire sweep:
Determines if the sweep is stopped immediately if a limit is exceeded; options are:
• Abort the sweep if the source limit is exceeded: ON (default)
• Complete the sweep if the source limit is exceeded:
Determines if the sweep runs from start to stop and then from stop to start:
• Sweep from start to stop only: OFF (default)
• Sweep from start to stop, then stop to start:
A string that indicates the reading buffer; the default buffers (defbuffer1 or
defbuffer2) or the name of a user-defined buffer; if no buffer is specified,
is used