Example script
The following example demonstrates how to connect to a remote device that is not TSP
®
enabled,
and send and receive data from this device:
-- Disconnect all existing TSP-Net connections.
tspnet.reset()
-- Set tspnet timeout to 5 s.
tspnet.timeout = 5
-- Establish connection to another device with IP address 192.168.1.51
-- at port 1394.
id_instr = tspnet.connect("192.168.1.51", 1394, "*rst\r\n")
-- Print the device ID from connect string.
print("ID is: ", id_instr)
-- Set the termination character to CRLF. You must do this
-- for each connection after the connection has been made.
tspnet.termination(id_instr, tspnet.TERM_CRLF)
-- Send the command string to the connected device.
tspnet.write(id_instr, "*idn?" .. "\r\n")
-- Read the data available, then print it.
print("instrument write/read returns: ", tspnet.read(id_instr))
-- Disconnect all existing TSP-Net sessions.
tspnet.reset()
Remote instrument events
If the Model 2461 is connected to a TSP-enabled instrument through TSP-Net, all events that occur
on the remote instrument are transferred to the event log of the Model 2461. The Model 2461
indicates events from the remote instrument by prefacing these events with “Remote Error.” For
example, if the remote instrument generates event code 4909, "Reading buffer not found within
device," the Model 2461 generates the string Remote Error: (4909) Reading buffer not found within
device.”
TSP-Net instrument commands: General device control
The following instrument commands provide general device control:
tspnet.clear() (on page 8-324)
tspnet.connect() (on page 8-325)
tspnet.disconnect() (on page 8-326)
tspnet.execute() (on page 8-327)
tspnet.idn() (on page 8-328)
tspnet.read() (on page 8-328)
tspnet.readavailable() (on page 8-329)
tspnet.reset() (on page 8-330)
tspnet.termination() (on page 8-331)
tspnet.timeout (on page 8-332)
tspnet.write() (on page 8-335)
TSP-Net instrument commands: TSP-enabled device control
The following instrument commands provide TSP-enabled device control:
tspnet.tsp.abort() (on page 8-332)
tspnet.tsp.abortonconnect (on page 8-333)
tspnet.tsp.rbtablecopy() (on page 8-334)
tspnet.tsp.runscript() (on page 8-335)

Table of Contents

Other manuals for Keithley 2461

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2461 and is the answer not in the manual?

Keithley 2461 Specifications

General IconGeneral
BrandKeithley
Model2461
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals